The evaluation of entrance surface dose and quality of radiograph in thorax radiological examination

Closed

S. Hermawan Dwi, Johan Noor, Y.P.W. Yuyun

2018 Journal of Global Pharma Technology Vol. 10 Issue 10 Article Cited by 0 Quartile

Abstract

Introduction: Computed Radiography (CR) is a technology of digitally medical imaging which is applied in x-ray generator. It is relatively easy to operate. Hence, the operators often use high parameter because they assume that it can ease image post-processing without considering radiation dose resulted by the process which is further received by patient’s organ penetrated by x-ray. It is mostly used on radiological examination of thorax at X-ray Department. Objective: This article aims to evaluate the relationship between ESD and the quality of resulted thorax radiograph as a protecting effort from radiation using ALARA principle. Method: The data were collected by taking pictures of Longman Phantom with kVp variations from 45 kVp to 105 kVp at 2 mAs to 32 mAs. TLD was utilized for the ESD measurement. Then, the radiographs were submitted to a radiologist to read and to convert them into diagnostic value. Result: This article results in several satisfactory imaging qualities based on the statistics of third quartile, that is on the radiation parameter of 75 kVp/16 mAs with dose value of 1.05 mSv, and of 85 kVp/8 mAs with dose value of 1.11 mSv, while BAPETEN allows parameter with dosage under 0.4 mSv on 55 kVp/8mAs, 65 kVp/4 mAs, 75 kVp/4 mAs, 85 kVp/2 mAs, and 105 kVp/2 mAs. Conclusion: The result of the research shows that setting the same value of kVp and increasing the value of mAs will increase ESD value. © 2009-2018, JGPT. All Rights Reserved.

Affiliations

Faculty of Mathematics and Natural Science, University of Brawijaya, Indonesia; Faculty of Medicine, University of Brawijaya, Indonesia